AC capacitor multi-parameter sorting system
Specializing in the production of capacitor measuring instrument series, inductance measuring instrument series, LCR digital bridge series, leakage current measuring instrument series, insulation resistance measuring instrument series, various capacitor multi-parameter automatic sorting system and special equipment
Classification:
Key words:
capacitor measuring instrument | inductance measuring instrument | capacitor
Product Description
1. Overview
This system is an automatic sorting system for AC capacitors. When in use, as long as the tested capacitor is inserted into the fixture of the system, the system can automatically test and sort different items of the tested parts according to different test requirements, and collect the test results of each item by classification (qualified products and unqualified products).
2. System Overview
The system is divided into "test discrimination" and "transmission sorting" two parts. The test fixture is installed on the test disc of the transmission sorting part. The operator will send the tested capacitor into the fixture of the system. The system will automatically transmit it to the test head of each test part for automatic test. According to the measured parameters, the qualified products or unqualified products of each grade will be classified and collected separately. (Centralized sorting of nonconforming products of this machine). The system can test and judge the following items of capacitors according to the process:
ACTV1→ACTV2→IR→CAP.DNG1→ΔC1
Technical specifications of 3. measuring instruments
3.1 AC withstand voltage (ACTV1.2) test
Test voltage: AC 0~1200V
Test times: 1 time
ACTV one station charging test
Current setting range: 0.1A ~ 3.8A
Unqualified detection: those whose measured value is greater than the set value are unqualified and will be automatically rejected after discharge displacement.
3.2 insulation resistance (IR) test
Test voltage: DC 50V, 100V, 250V, 500V
Measurement range: 100MΩ ~ 990MΩ
1000MΩ ~ 9900MΩ
10000MΩ ~ 99000MΩ
Setting: in the test range of two-digit arbitrary setting
Test accuracy: 102MΩ is ± 10%,103MΩ is ± 10%,104MΩ is ± 20%.
Unqualified detection: the capacitor whose measured value is lower than the set value is automatically rejected after discharging three stations.
3.3 capacity error (△C) test and loss tangent (DNG) measurement and sorting
Please refer to the TE2616A instruction manual for detailed instructions.
4. other technical specifications
4.1 power supply
The measuring instrument part is single-phase AC220V ± 10% 50Hz
Transmission part is three-phase AC 380V ± 10% 50Hz
AC cabinet is three-phase AC 380V ± 10% 50Hz 8kW
4.2 Test Fixture
The total number of fixtures is 24.
4.3 belt drive and variable speed
Drive belt speed Max:2S/station step continuously adjustable, controlled by frequency converter.
4.4 fault protection function
4.4.1 Air pressure protection: If the air pressure is insufficient, an audible and visual alarm will be issued after starting up, and starting will be prohibited until the fault is eliminated;
4.4.2 Magnetic sensitive switch protection: if the initial position of NG, GOOD blanking cylinder or test cylinder is not in place or during operation, an audible and visual alarm will be issued and start-up will be prohibited until the fault is eliminated;
4.4.3 Overload protection: the mechanical transmission is protected by torque limiter and proximity switch. Overload audible and visual alarm.
All faults are automatically displayed on the touch screen. After troubleshooting, press the reset key on the touch screen or the stop button of the control box to reset and restart.
4.5 environmental requirements
Temperature: 5 ℃ ~ 35 ℃
Humidity: <75%